Resumen

Background: While robotic-arm assisted total knee arthroplasty (RA-TKA) has seen a major increase in its utilization, it requires bone array pins to be fixed into the femur and tibia, which intrinsically carries a risk. As it is currently off-label with some robotic platforms to place pins intraincisional, we aimed to evaluate the safety of intraincisional pin placement during RA-TKAs. Methods: A prospective cohort of 2,343 patients who underwent RA-TKA at a North American Healthcare System between January 2018 and March 2022 was included. Primary outcomes included periprosthetic fracture or infection (eg, superficial or deep). Secondary outcomes included 1-year reoperation rate due to any cause. Cases were retrospectively reviewed to determine whether complications could be attributed to metaphyseal intraincisional pin placement (4.0 mm pins; two tibial and two femoral). The 90-day follow-up was 100% and the 1-year follow-up rate was 70.6% (n = 1,655). Results: The pin-site related periprosthetic fracture incidence at 90 days was 0.09% (2 out of 2,343). The 90-day infection incidence was 1.4% (superficial: 22; deep: 13). The 1-year reoperation rate was 1.8% (29 out of 1,655). The most common causes of reoperation at 1-year were deep infection (n = 14; 0.83%), superficial infection (n = 3; 0.18%), periprosthetic fracture, mechanical symptoms, instability, and hematoma (n = 2; 0.12% for each). Conclusions: One in 1,172 patients may experience a pin-related periprosthetic fracture after RA-TKA with intraincisional bone array pin placement. There was a low 90-day infection incidence and reoperations within 1-year after RA-TKA were rare.

Idioma originalInglés
Páginas (desde-hasta)910-915.e1
PublicaciónJournal of Arthroplasty
Volumen39
N.º4
DOI
EstadoPublicada - abr. 2024
Publicado de forma externa

Huella

Profundice en los temas de investigación de 'Intraincisional Pin Placement is Safe for Robotic-Assisted Total Knee Arthroplasty'. En conjunto forman una huella única.

Citar esto