HPGe detectors long time behaviour in high-resolution γ spectrometry

L. Sajo-Bohus, D. Rosso, A. M. Sajo Castelli, D. R. Napoli, E. Fioretto, R. Menegazzo, H. Barros, C. A. Ur, D. Palacios, J. Liendo

Producción científica: Contribución a una revistaArtículorevisión exhaustiva

11 Citas (Scopus)

Resumen

A large set of data on long term performance of n-type HPGe detectors used in GASP, EUROBALL and CLARA γ spectrometers, as well as environmental measurements have been collected over two decades. In this paper a detailed statistical analysis of this data is given and detector long term behaviour is provided to the scientific community. We include failure, failure mode, repair frequency, repair outcome and its influence in the energy efficiency and energy resolution. A remarkable result is that the life span distribution is exponential. A detector's failure is a memory-less process, where a previous failure does not influence the upcoming one. Repaired spectrometers result in high reliability with deep implications in the management of large scale high-resolution gamma spectrometry related projects. Findings show that on average, detectors initial counting efficiency is slightly lower (∼2%) than that reported by the manufacturers and the repair process (including annealing) does not affect significantly the energy efficiency, even after a long period of use. Repaired detector energy resolution statistics show that the probability, that a repaired detector will be at least as good as it was originally, is more than 3/4.

Idioma originalInglés
Páginas (desde-hasta)132-138
Número de páginas7
PublicaciónNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Volumen648
N.º1
DOI
EstadoPublicada - 21 ago. 2011
Publicado de forma externa

Huella

Profundice en los temas de investigación de 'HPGe detectors long time behaviour in high-resolution γ spectrometry'. En conjunto forman una huella única.

Citar esto