Masi: a mechanical ventilator based on a manual resuscitator with telemedicine capabilities for patients with ARDS during the COVID-19 crisis

Javier Chang · A P Acosta · Jorge Benavides-Aspiazu · Jaime Reategui · Christiam Rojas · Jordi Cook · Richard Nole · Luigi Giampietri · Sandra Pérez-Buitrago · Fanny L. Casado

Resumen

In this article, we introduce a portable and low-cost ventilator that could be rapidly manufactured, to meet the increasing demand of ventilators worldwide produced by COVID-19 pandemic. These ventilators should be rapidly deployable and with functional capabilities to manage COVID-19 patients with severe acute respiratory distress syndrome (ARDS). Our implementation offers robustness, safety and functionality absent in existing solutions to the ventilator shortage (i.e., telemonitoring, easy-to-disinfect, modularity) by maintaining simplicity. The design makes use of a manual resuscitator as the core respiration component activated by a compression mechanism which consist of two electronically controlled paddles. The quality measurements obtained after testing on a calibrated artificial lung demonstrate repeatability and accuracy exceeding human capabilities of manual ventilation. The complete design files are provided in the supplementary materials to facilitate ventilator production even in resource-limited settings. The implementation of this mechanical ventilator could eliminate device rationing or splitting to serve multiple patients on ICUs.

Autores y colaboradores

Authors

Javier Chang
A P Acosta
Jorge Benavides-Aspiazu
Jaime Reategui
Christiam Rojas
Jordi Cook
Richard Nole
Luigi Giampietri
Sandra Pérez-Buitrago
Fanny L. Casado

Palabras clave

Acute respiratory distress syndrome COVID-19 pandemic Critical care Mechanical ventilation Respiratory insufficiency