TY - GEN
T1 - Performance of the Masi Peruvian ventilator at high altitude
AU - Perez-Buitrago, Sandra
AU - Gomez-Alzate, Daniela
AU - Cordova, Mauricio
AU - Rojas, Christiam
AU - Chang, Javier
AU - Castaneda, Benjamin
N1 - Publisher Copyright:
© 2021 IEEE.
PY - 2021
Y1 - 2021
N2 - In response to Covid-19 crisis, 310 Masi ventilators were produced and validated in Lima, Peru, according to applicable standards. Four of them, were transported to Puno, in order to strengthen ICU Services there, but this set a major challenge to Masi team as effects of altitude on ventilators were unknown. Once there, ventilators were acclimated and calibrated. Volume tidal, I:E ratio, respiratory frequency and PEEP were tested, all of them presenting errors under 15%, except for tidal volume, for which a 25% negative correction was applied. After the installation of a new version of Masi software, parameters were tested again, all of them presenting results with errors below 15%, which allowed the Masi team to take them to ICU services for use.Clinical Relevance - Masi Peruvian Ventilators are able to perform according to their specifications at extremely high altitude, after the adequate calibration. These devices are an alternative to treat COVID-19 patients in the middle of the crisis.
AB - In response to Covid-19 crisis, 310 Masi ventilators were produced and validated in Lima, Peru, according to applicable standards. Four of them, were transported to Puno, in order to strengthen ICU Services there, but this set a major challenge to Masi team as effects of altitude on ventilators were unknown. Once there, ventilators were acclimated and calibrated. Volume tidal, I:E ratio, respiratory frequency and PEEP were tested, all of them presenting errors under 15%, except for tidal volume, for which a 25% negative correction was applied. After the installation of a new version of Masi software, parameters were tested again, all of them presenting results with errors below 15%, which allowed the Masi team to take them to ICU services for use.Clinical Relevance - Masi Peruvian Ventilators are able to perform according to their specifications at extremely high altitude, after the adequate calibration. These devices are an alternative to treat COVID-19 patients in the middle of the crisis.
UR - http://www.scopus.com/inward/record.url?scp=85122497794&partnerID=8YFLogxK
U2 - 10.1109/EMBC46164.2021.9630947
DO - 10.1109/EMBC46164.2021.9630947
M3 - Conference contribution
C2 - 34892337
AN - SCOPUS:85122497794
T3 - Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS
SP - 5031
EP - 5034
BT - 43rd Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2021
PB - Institute of Electrical and Electronics Engineers Inc.
T2 - 43rd Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2021
Y2 - 1 November 2021 through 5 November 2021